12 |
![]() |
2005
(a)
Set CALIBRATE potentiometer R39 to
the middle of
its range.
(b)
Disconnect any jumper
wires connected across re-
sistors R9, R13, R40 and/or R41.
(c)
Connect jumper wires, in various combinations,
across resistors R9, R13, R40, and R41 until the
output is within 2.5 MV of zero. Solder the
jumper wires in these positions.
(d)
Adjust CALIBRATE potentiometer R39 for an
output
of 0.000 volt
±1 MV.
c.
20—Volt Adjustment:
NOTE
Make this adjustment only after the zero
voltage calibration.
(1)
Set the RANGE switch to 10—20V and
adjust the supply
output to 20.000 volts. Set the VERNIER control fully counterclockwise.
(2)
Check that the dots on the VERNIER control and
front
panel are aligned. A setscrew is located in the VERNIER control for
any
necessary adjustment.
(3)
Set the
VERNIER control to 0.
(4)
Connect a high precision voltmeter across the output
of the
supply.
(5)
Sot the AC switch to ON and observe the voltmeter.
(6)
If necessary, adjust potentiometer R12 (on the amplifier
board) until the voltmeter reads 20.000 ±0.007
volts.
d.
CURRENT LIMIT ADJ Range Adjustment:
(1)
Set the power supply output voltage to 20.000 volts.
(2)
Sot the meter switch to MA.
(3)
Set potentiometer R27 (on the amplifier board)
to
the
center
of its range.
(4)
Adjust CURRENT LIMIT ADJ potentiometer R25 through its
entire range while depresseing CURRENT LIMIT SET pushbutton S3.
(5)
Adjust potentiometer R27 until the CURRENT LIMIT ADJ
potentiometer varies the output current from 0
to
500 MA over its
entire range.
645
4—2
|
![]() |
![]() |
12 |
![]() |