OUTPUT
—
0 to 20 VDC, continuously adjustable, at 0 to 500 MA.
INPUT
—
105 to 125 V, 50 to 440 Hz, 40 W max. (including oven heater).
CALIBRATION AND ACCURACY
—
Two dual, concentric, decade switches provide dial readout to four places with an accuracy of 0.1%
±
1 MV at outputs below 10 V. A 1-MV range, single-turn potentiometer permits interpolation of the last place. This potentiometer has a resolution of 10
uV.
A toggle switch extends the range of the dial readout from 10 to 20 V, without affecting resolution or accuracy (five-place readout). An illuminated indicator identifies the range in operation.
REGULATION
—
Output voltage change less than 100
uV
(at sense connection points) for line or load variations over the operating range.
RIPPLE AND NOISE
—
Less than 100
uV
peak-to-peak.
SOURCE IMPEDANCE
—
Less than 0.2 milliohm at DC, 0.04 ohm at 20 KHz, 0.5 ohm at 1 MHz.
RECOVERY TIME
—
Less than 10 uS to return to within 250
uV
or 0.005% (whichever is greater) of the set voltage for a 10% to 100% step change in rated load; less than 40
uS
to return with 100
uV.
OPERATING TEMPERATURE RANGE
—
15 to 4 °C ambient; 0 to 60 °C ambient with reduced accuracy.
TEMPERATURE COEFFICIENT
—
DC output voltage change
less than 0.001% or 50
uV
(whichever is greater) per
°C over the range of
+
15 to +45 °C; less than 0.002%
or 100
uV
(whichever is greater) per °C from 0 to + 15
°C and +45 to +60 °C.
STABILITY
—
Better than 0.001%
+
100 MV/8 hrs.; better than 1 MV/wk (at constant ambient temperature, load and line voltage, after warmup).
|
OUTPUT TERMINALS
—
Three front-panel binding posts and rear-panel barrier block. Either positive or negative output terminal may be grounded.
METER
—
Front-panel volt-ammeter permits monitoring output voltage or current with an accuracy of ±2%.
REMOTE SENSING
—
Two terminals on rear-panel barrier block are provided for remote sensing of voltage at the load.
REMOTE PROGRAMMING
—
Rear terminals are provided for remote programming of the output voltage at
1000
ohms-per-volt. Accuracy of programming is better than 0.01% of value of resistance (including connecting leads).
CURRENT LIMITING
—
The output current can be limited to any value from 0 to 500 MA by a front-panel control. A push-button permits easy adjustment without the need for shorting the output terminals.
DIMENSIONS
—
8 3/8”
x 4 3/4" x
8-15/16”
WEIGHT
—
9
pounds
FINISH
—
Brushed anodized aluminum panel with etched black lettering. Dust cover finished in blue enamel; chassis and bottom plate, golden indite.
|
![]() |
![]() |